The first experiment on generation of high-order harmonics with L1

This experiment was a culmination of a year-long installation of the L1-ALLEGRA laser, the L1-E1 beam transport, and of the HHG beamline, which were integrated into one system for user experiments.

In this first integrated test experiment, the ALLEGRA  laser front-end output was propagated through the whole system and compressed in vacuum by chirped mirrors resulting in the following laser output parameters: Pulse energy: 1.4 mJ, pulse duration: 14 fs, central wavelength: 830 nm, repetition rate: 1 kHz. The beam was then propagated though the E1 vacuum beam transport delivering it to the HHG beamline. There it was characterized and focused into a 2 cm long gas cell. Good pointing stability was measured, which is a crucial feature for pump-probe experiments with tight focusing.

The harmonic generation performance of L1 was compared to the previously available commercial laser driver (Coherent Astrella: 5 mJ, 40 fs, 1 kHz) used in E1. The L1 driven spectrum was found to have higher efficiency below 30 nm and generate shorter wavelengths (see Fig.2).

The experimental team will continue research and development of the kHz X–ray sources in E1, namely the plasma X-ray source, which generates 10 keV radiation in sub-picosecond pulses.

 

Fig. 1 HHG Beamline with a raw image from an XUV spectrometer showing the high order harmonics generated by L1 laser pulses. Members of the HHG group of the RP2 team from left Ondřej Finke, Jaroslav Nejdl, and Ondřej Hort.

Fig. 2. Spectrum of the high order harmonics generated in gas cell filled with argon driven by a commercial laser driver with 40 fs-long pulses (blue) and with L1 laser pulses (red). The harmonics cutoff in logarithmic scale is shown in the inset.